DR 600 - SB291 - Spellman Generator ZST2 FW V3.11r28 with SPM 5.20 solves too high dose values of IBA DAP

Published: Wed Jul 16 15:21:48 2025
Created Date: July 10, 2025
Document Type:
Service Bulletin
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR 600
Item:
DR 600
Summary:
DR 600 - Service Bulletin No. 291 - Version 1.0
- In the last months several cases have been reported of sporadically too high displayed dose values on systems with an IBA DAP chamber.
- To solve this problem an updated Spellman Generator ZST2 board with Firmware V3.11r28 has been released.
- The new ZST2 Firmware V3.11r28 requires at least a new SPM software version 5.20.

Affected systems: Replace the ZST2 board by ZST2 V3.11r28 on systems where the customer reports the problem.
Publication History:
Published
2025-07-16 15:21:48 CET
Additional Languages:
No additional languages found

X-Ray Generator EDITOR HFe - Software - SPM 5.20

Published: Wed Jul 16 15:21:46 2025
Created Date: July 02, 2025
Document Type:
Software
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR Components
Item:
Generator
Summary:
Generator_SW_Spellman_SPM_5.20.0100.zip
This software is required for ZST2 FW V3.11r28 - For details refer to DR 600 - SB291, Document ID 85203167.
Software Download via AgfaBox directory (Agfa ID and password required):
Connect to https://agfabox.agfa.net/ and navigate to the file:
Radiology Solutions SW > Imaging_Software > Direct_Radiography > Generator > Spellman
Publication History:
Published
2025-07-16 15:21:46 CET
Additional Languages:
No additional languages found

Valory - SB076 - Spellman Generator ZST2 FW V3.11r28 with SPM 5.20 solves too high dose values of IBA DAP

Published: Wed Jul 16 15:13:41 2025
Created Date: July 16, 2025
Document Type:
Service Bulletin
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
Valory
Item:
Valory
Summary:
Valory - Service Bulletin No. 76 - Version 1.0
- In the last months several cases have been reported of sporadically too high displayed dose values on systems with an IBA DAP chamber.
- To solve this problem an updated Spellman Generator ZST2 board with Firmware V3.11r28 has been released.
- The new ZST2 Firmware V3.11r28 requires at least a new SPM software version 5.20.

Affected systems: Replace the ZST2 board by ZST2 V3.11r28 on systems where the customer reports the problem.
Publication History:
Published
2025-07-16 15:13:41 CET
Additional Languages:
No additional languages found

DR 400 - SB228 - Spellman Generator ZST2 FW V3.11r28 with SPM 5.20 solves too high dose values of IBA DAP

Published: Wed Jul 16 15:12:34 2025
Created Date: July 16, 2025
Document Type:
Service Bulletin
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR 400
Item:
DR 400
Summary:
DR 400 - Service Bulletin No. 228 - Version 1.0
- In the last months several cases have been reported of sporadically too high displayed dose values on systems with an IBA DAP chamber.
- To solve this problem an updated Spellman Generator ZST2 board with Firmware V3.11r28 has been released.
- The new ZST2 Firmware V3.11r28 requires at least a new SPM software version 5.20.

Affected systems: Replace the ZST2 board by ZST2 V3.11r28 on systems where the customer reports the problem.
Publication History:
Published
2025-07-16 15:12:34 CET
Additional Languages:
No additional languages found

Peissenberg Radiology Solutions Services - Classroom Training Schedule

Published: Tue Jul 15 09:49:28 2025
Created Date: July 11, 2025
Document Type:
Service Training
Availability:
Extranet
Category:
General Info
Language:
English
Subcategory:
Service
Item:
All Products
Summary:
Peißenberg Radiology Solutions Services - Classroom Training Schedule
Q3-2025
Version: 2025-07-1
NOTE: For using the course links in the pdf file you need to login to MyLearning first (MyLearning licence required).
Publication History:
Published
2025-07-15 09:49:28 CET
Published
2025-07-10 13:38:15 CET
Published
2025-06-13 15:57:49 CET
Published
2025-06-03 17:08:21 CET
Published
2025-05-20 09:27:32 CET
Published
2025-04-30 12:06:29 CET
Published
2025-04-15 10:29:34 CET
Published
2025-02-20 09:25:09 CET
Published
2025-02-19 14:55:55 CET
Published
2025-02-11 15:54:49 CET
Published
2025-02-06 09:38:16 CET
Published
2025-01-23 14:11:49 CET
Published
2025-01-13 10:51:25 CET
Published
2024-11-12 08:38:50 CET
Published
2024-10-24 09:00:20 CET
Published
2024-10-01 12:01:20 CET
Published
2024-09-24 08:34:01 CET
Published
2024-09-13 10:38:12 CET
Published
2024-09-13 10:35:16 CET
Published
2024-08-14 14:16:52 CET
Published
2024-07-15 08:07:36 CET
Published
2024-07-09 14:38:35 CET
Published
2024-06-13 14:36:57 CET
Published
2024-05-29 10:43:27 CET
Published
2024-04-15 11:49:59 CET
Published
2024-04-11 11:46:12 CET
Published
2024-04-03 16:34:45 CET
Published
2024-04-03 12:56:49 CET
Published
2024-03-28 15:52:56 CET
Published
2024-03-13 14:31:28 CET
Published
2024-03-11 12:00:04 CET
Published
2024-03-04 08:00:38 CET
Published
2024-01-22 11:33:48 CET
Published
2023-11-09 07:55:24 CET
Published
2023-10-26 09:39:03 CET
Published
2023-10-13 07:58:18 CET
Published
2023-10-10 09:16:45 CET
Published
2023-09-13 08:16:40 CET
Published
2023-08-23 12:32:30 CET
Published
2023-08-10 16:19:47 CET
Published
2023-07-26 13:43:16 CET
Published
2023-06-21 08:43:31 CET
Published
2023-05-09 09:07:20 CET
Published
2023-04-18 10:49:24 CET
Published
2023-04-12 11:57:59 CET
Published
2023-03-28 11:04:39 CET
Published
2023-03-27 07:55:11 CET
Published
2023-02-20 12:04:59 CET
Published
2023-02-16 14:33:07 CET
Published
2023-02-07 12:25:15 CET
Published
2023-01-25 14:23:32 CET
Published
2023-01-17 16:32:31 CET
Published
2023-01-13 14:19:36 CET
Published
2022-11-29 12:26:49 CET
Published
2022-11-21 14:19:33 CET
Published
2022-11-14 13:25:26 CET
Published
2022-11-10 17:07:56 CET
Published
2022-10-13 15:53:37 CET
Published
2022-10-07 09:29:23 CET
Published
2022-09-15 15:41:57 CET
Published
2022-08-22 13:22:00 CET
Published
2022-08-12 08:11:25 CET
Published
2022-07-29 08:06:33 CET
Published
2022-07-14 11:33:05 CET
Published
2022-05-30 14:17:01 CET
Published
2022-05-30 07:46:38 CET
Published
2022-05-09 11:21:29 CET
Published
2022-04-25 14:36:31 CET
Published
2022-04-13 10:43:07 CET
Published
2022-04-07 10:19:48 CET
Published
2022-03-30 08:53:39 CET
Published
2022-03-28 08:00:15 CET
Published
2022-03-22 07:31:33 CET
Published
2022-02-23 10:22:27 CET
Published
2022-02-16 09:50:26 CET
Published
2022-02-09 14:55:49 CET
Published
2022-01-26 07:28:14 CET
Published
2022-01-18 14:23:54 CET
Published
2021-12-22 09:03:54 CET
Published
2021-12-17 07:09:37 CET
Published
2021-11-25 15:06:24 CET
Published
2021-10-20 12:57:46 CET
Published
2021-10-15 14:48:58 CET
Published
2021-09-30 07:17:22 CET
Published
2021-09-29 09:48:57 CET
Published
2021-09-22 14:56:09 CET
Published
2021-09-08 06:48:51 CET
Published
2021-08-17 11:04:16 CET
Published
2021-08-11 10:11:13 CET
Published
2021-07-28 08:34:27 CET
Published
2021-07-20 12:16:48 CET
Published
2021-07-15 07:21:40 CET
Published
2021-07-07 10:57:46 CET
Published
2021-06-25 06:41:08 CET
Published
2021-06-01 13:25:58 CET
Published
2021-05-05 10:31:58 CET
Published
2021-04-28 10:36:23 CET
Published
2021-04-12 13:47:11 CET
Published
2021-03-30 14:27:34 CET
Published
2021-03-16 07:49:10 CET
Published
2021-03-09 08:28:36 CET
Published
2021-03-02 09:22:34 CET
Published
2021-03-02 09:22:31 CET
Published
2021-02-15 10:12:58 CET
Published
2021-01-26 08:49:31 CET
Published
2021-01-18 07:54:37 CET
Published
2020-12-15 14:13:11 CET
Published
2020-12-03 08:30:55 CET
Published
2020-11-30 15:03:47 CET
Published
2020-11-18 09:10:24 CET
Published
2020-10-26 14:36:20 CET
Published
2020-10-14 11:27:40 CET
Published
2020-09-01 11:23:35 CET
Published
2020-08-25 08:47:35 CET
Published
2020-08-12 14:33:22 CET
Published
2020-08-04 07:52:24 CET
Published
2020-07-22 15:06:46 CET
Published
2020-07-20 10:03:09 CET
Published
2020-07-06 14:17:12 CET
Published
2020-06-23 13:02:26 CET
Published
2020-05-20 07:22:58 CET
Published
2020-05-06 07:22:10 CET
Published
2020-04-28 10:52:03 CET
Published
2020-04-16 14:37:54 CET
Published
2020-04-07 09:37:31 CET
Published
2020-03-31 10:59:13 CET
Published
2020-03-27 11:14:30 CET
Published
2020-03-23 08:32:00 CET
Published
2020-03-17 08:22:01 CET
Published
2020-03-16 14:20:13 CET
Published
2020-02-03 14:02:48 CET
Published
2020-01-23 11:51:59 CET
Published
2020-01-20 14:20:33 CET
Published
2020-01-08 10:54:40 CET
Published
2019-12-17 11:09:08 CET
Published
2019-12-04 07:55:18 CET
Published
2019-11-27 13:21:02 CET
Published
2019-11-21 12:09:25 CET
Published
2019-11-11 07:49:22 CET
Published
2019-10-16 09:28:33 CET
Published
2019-09-20 08:06:30 CET
Published
2019-07-24 14:55:25 CET
Published
2019-07-02 11:37:59 CET
Published
2019-06-07 13:28:08 CET
Published
2019-06-04 07:31:37 CET
Published
2019-05-21 09:18:13 CET
Published
2019-05-17 10:56:45 CET
Published
2019-05-10 15:49:39 CET
Published
2019-05-06 09:36:57 CET
Published
2019-04-04 16:00:18 CET
Published
2019-03-07 08:25:02 CET
Published
2019-02-12 15:42:21 CET
Published
2019-02-01 07:57:21 CET
Published
2019-01-21 12:50:27 CET
Published
2019-01-11 14:37:57 CET
Published
2018-10-26 16:37:09 CET
Published
2018-10-11 13:46:25 CET
Published
2018-10-05 13:30:00 CET
Published
2018-08-27 16:06:12 CET
Published
2018-08-20 14:04:00 CET
Published
2018-07-10 08:09:37 CET
Published
2018-07-09 10:08:48 CET
Published
2018-07-03 15:08:11 CET
Published
2018-06-11 15:31:25 CET
Published
2018-05-30 08:30:32 CET
Published
2018-05-22 08:20:43 CET
Published
2018-04-25 14:21:01 CET
Published
2018-04-16 13:03:02 CET
Published
2018-04-05 16:06:42 CET
Published
2018-02-13 08:59:20 CET
Published
2018-02-05 13:40:37 CET
Published
2018-01-15 15:24:08 CET
Published
2017-12-07 10:20:09 CET
Published
2017-10-24 15:31:49 CET
Published
2017-08-01 13:16:58 CET
Published
2017-04-11 15:54:26 CET
Published
2017-03-30 13:17:47 CET
Published
2016-11-29 08:15:49 CET
Published
2016-11-10 13:07:58 CET
Published
2016-10-24 11:28:55 CET
Published
2016-09-29 14:47:27 CET
Published
2016-07-07 11:44:32 CET
Published
2016-05-12 11:18:48 CET
Published
2016-04-12 09:23:48 CET
Published
2016-03-14 16:56:15 CET
Published
2016-02-01 10:07:23 CET
Published
2015-12-04 11:23:02 CET
Published
2015-11-04 08:07:20 CET
Published
2015-08-12 08:11:07 CET
Published
2015-07-08 10:53:51 CET
Published
2015-06-03 14:58:41 CET
Published
2015-05-07 13:06:00 CET
Published
2015-03-03 12:47:33 CET
Published
2015-01-15 08:30:23 CET
Published
2014-11-18 16:24:58 CET
Published
2014-09-18 13:37:08 CET
Published
2014-09-04 08:43:35 CET
Additional Languages:
No additional languages found

MUSICA Workstation Patient Care Packages (Datasheet_English)

Published: Mon Jul 14 14:48:15 2025
Created Date: July 14, 2025
Document Type:
Data Sheet
Availability:
Internet
Category:
Agfa Workstation
Language:
English
Subcategory:
NX Software
Item:
NX 27.00
Summary:
Publication History:
Published
2025-07-14 14:48:15 CET
Additional Languages:

MUSICA Workstation Productivity Packages (Datasheet_English)

Published: Mon Jul 14 14:38:11 2025
Created Date: July 08, 2025
Document Type:
Data Sheet
Availability:
Internet
Category:
Agfa Workstation
Language:
English
Subcategory:
NX Software
Item:
NX 27.00
Summary:
Publication History:
Published
2025-07-14 14:38:11 CET
Additional Languages:

MUSICA Workstation Image Quality Packages (Datasheet_English)

Published: Mon Jul 14 14:26:06 2025
Created Date: July 08, 2025
Document Type:
Data Sheet
Availability:
Internet
Category:
Agfa Workstation
Language:
English
Subcategory:
NX Software
Item:
NX 27.00
Summary:
Publication History:
Published
2025-07-14 14:26:06 CET
Additional Languages:

MUSICA Workstation Data Protection Packages (Datasheet_English)

Published: Mon Jul 14 14:23:02 2025
Created Date: July 08, 2025
Document Type:
Data Sheet
Availability:
Internet
Category:
Agfa Workstation
Language:
English
Subcategory:
NX Software
Item:
NX 27.00
Summary:
Publication History:
Published
2025-07-14 14:23:02 CET
Additional Languages:

DR 800 - Table GMM - Service Manual

Published: Mon Jul 14 11:13:22 2025
Created Date: July 11, 2025
Document Type:
Service Manual
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR 800
Item:
DR 800
Summary:
DR 800 Table - Service Manual, only Type 6010/200
Edition 1, Revision 15
Changes compared to previous revision: Updated Installation Manual, Rev. 14, April 2025

This Service Manual comprises the following documents:
- DR 800 Table Installation Manual, CODE 62347 - REVISION 14, April 2025
- DR 800 Table Adjustment Manual, CODE 62361 - REVISION 04, September 2023
- DR 800 Table Troubleshooting, CODE 62373 - REVISION 04, March 2024
- DR 800 Table Schemes, CODE 62356-AGF - REVISION 03, April 2020
- DR 800 Terminal service software V 2.3 reference manual, CODE 62259 - Vers. 2.3, January 2017
- DR 800 Preventive Maintenance tables, Vers. 05, September 2022
Publication History:
Published
2025-07-14 11:13:22 CET
Published
2025-04-17 08:24:58 CET
Published
2024-08-01 08:44:53 CET
Published
2024-04-18 10:25:44 CET
Published
2023-12-05 08:43:23 CET
Published
2022-04-20 09:31:00 CET
Published
2021-06-07 12:38:43 CET
Published
2021-03-31 14:27:05 CET
Published
2020-12-15 14:41:28 CET
Published
2020-07-22 13:54:25 CET
Published
2020-04-08 14:56:38 CET
Published
2019-08-06 15:27:57 CET
Published
2019-04-11 16:32:09 CET
Published
2019-03-25 14:43:54 CET
Published
2018-12-13 14:55:37 CET
Published
2018-11-15 09:09:33 CET
Published
2018-07-27 15:25:57 CET
Published
2018-06-21 15:11:26 CET
Additional Languages:
No additional languages found

Enclosure - DR 100e - NX Clean Installation Procedure - CyberMed PC

Published: Fri Jul 11 08:57:07 2025
Created Date: July 09, 2025
Document Type:
Service Enclosures
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR 100e
Item:
DR 100e
Summary:
NX Workstation: NX Clean Installation Procedure for All in One PC CyberMed NB19
The zip file contains NX-specific clean installation procedures:
- DR 100e - NX_Clean_Installation_Procedure - CyberMed_PC-NX 21.00SU1.pdf
- DR 100e - NX_Clean_Installation_Procedure - CyberMed_PC-NX 22.00SU1.pdf
- DR 100e - NX_Clean_Installation_Procedure - CyberMed_PC-NX 23.00SU1.pdf
- DR 100e - NX_Clean_Installation_Procedure - CyberMed_PC-NX 24.00.pdf
- DR 100e - NX_Clean_Installation_Procedure - CyberMed_PC-NX 25.00.pdf
- DR 100e - NX_Clean_Installation_Procedure - CyberMed_PC-NX 26.00SU1.pdf
Publication History:
Published
2025-07-11 08:57:07 CET
Published
2024-07-08 10:54:49 CET
Published
2024-02-21 08:27:09 CET
Published
2021-11-24 15:10:16 CET
Published
2021-08-30 10:19:44 CET
Published
2021-08-18 12:17:17 CET
Published
2021-02-25 14:54:52 CET
Published
2020-02-05 15:18:47 CET
Published
2019-11-07 15:35:04 CET
Additional Languages:
No additional languages found

DR 100e - SB085 - CleanInstall set for NX 26 SU1 and NX 24 for AiO PC CyberMed NB19 available

Published: Fri Jul 11 08:55:56 2025
Created Date: June 30, 2025
Document Type:
Service Bulletin
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR 100e
Item:
DR 100e
Summary:
DR 100e - Service Bulletin No. 85 - Version 1.0
- It announces the DR100E CLEAN INST. CYB. NB19 NX26-SU1 SET for the DR 100e (digital, subtypes 003/004) AiO PC.
-The delivery of this set contains:
-- 1 x USB flash drive for clean installation of NX26.00 SU1 for WIN10 LTSC license
-- 1 x USB flash drive for clean installation of NX24.00 for WIN10 LTSB license
Affected systems:
All DR 100e with Windows 10 and a LTSC or a LTSB license, that have to be installed from scratch.
Windows 10 LTSC is installed as of SN 30172 (type 6012/003) and as of SN 40373 (type 6012/004). NX 26.00 SU1 is installed on DR 100e digital systems as of SN 30231 (type 6012/003) and as of SN 40445 (type 6012/004).
Publication History:
Published
2025-07-11 08:55:56 CET
Additional Languages:
No additional languages found

NX/MUSICA Acquisition Workstation - Software - Windows 10 Hotfixes 10.XXXX.047 (file size approx. 6.3 GB)

Published: Thu Jul 10 12:02:01 2025
Created Date: July 10, 2025
Document Type:
Software
Availability:
Extranet
Category:
Agfa Workstation
Language:
English
Subcategory:
NX Software
Item:
All NX versions
Summary:
HotFixes 10.XXXX.047.zip
Zip file contains:
- HotFixes 10.21H2.047.exe
- HotFixes 10.1607.047.exe
- HotFixes 10.1809.047.exe
- HotFixes 11.24H2.047.exe
- HotFix_Release_Note_10.21H2.047.pdf
- HotFix_Release_Note_10.1607.047.pdf
- HotFix_Release_Note_10.1809.047.pdf
- HotFix_Release_Note_11.24H2.047.pdf
- readme.txt
Note that Windows 11 is not released yet on NX PCs.
These are the same Windows 10 hotfixes for NX that can be replicated with FUSE or downloaded from ELMS.

This file can be downloaded from the AgfaBox (Agfa ID and password required).
Connect to https://agfabox.agfa.net/ and navigate to the file:
Radiology Solutions SW > Imaging_Software > Agfa_Workstation > NX Software >All NX versions > NX Hotfixes

In case of download problems: Download the split 7zip files which are available in "NX Windows 10 Hotfixes split in 1 GB files in case of download problems" directory.
They need to be combined with 7zip software after download. For details refer to readme_combining single 7ZIP files.txt.
Publication History:
Published
2025-07-10 12:02:01 CET
Published
2025-04-01 15:19:06 CET
Published
2024-12-17 11:40:47 CET
Published
2024-10-18 08:53:16 CET
Published
2024-06-20 16:22:58 CET
Published
2024-05-08 16:24:10 CET
Additional Languages:
No additional languages found

DR 100e with Fixed Column - Service Manual

Published: Thu Jul 10 10:50:25 2025
Created Date: March 18, 2025
Document Type:
Service Manual
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR 100e
Item:
DR 100e
Summary:
DR 100e with Fixed Column - Service Manual
Edition 1, Revision 9
Changes compared to previous revision:
- Added XF+ 10, XF+ 14, XF+ 17 detectors as new system components.
- Updated references.
Publication History:
Published
2025-07-10 10:50:25 CET
Published
2024-09-27 14:01:48 CET
Published
2024-07-08 10:52:33 CET
Published
2023-06-01 15:40:42 CET
Published
2022-06-29 16:59:57 CET
Published
2020-01-30 12:46:45 CET
Published
2019-11-07 14:57:57 CET
Published
2017-09-01 14:46:00 CET
Published
2017-04-12 13:08:08 CET
Additional Languages:
No additional languages found

DR 100e with Rotating Column - Service Manual

Published: Thu Jul 10 10:50:08 2025
Created Date: March 18, 2025
Document Type:
Service Manual
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR 100e
Item:
DR 100e
Summary:
DR 100e with Rotating Column - Service Manual
Edition 1, Revision 9
Changes compared to previous revision:
- Added XF+ 10, XF+ 14, XF+ 17 detectors as new system components.
- Updated references.
Publication History:
Published
2025-07-10 10:50:08 CET
Published
2024-09-27 14:02:00 CET
Published
2024-07-08 10:53:41 CET
Published
2023-06-01 15:40:29 CET
Published
2022-06-29 15:40:37 CET
Published
2020-01-30 12:46:35 CET
Published
2019-11-07 14:58:05 CET
Published
2017-09-01 14:45:46 CET
Published
2017-04-12 13:06:53 CET
Additional Languages:
No additional languages found

DR 100e - Solution Service Manual

Published: Thu Jul 10 10:48:52 2025
Created Date: April 14, 2025
Document Type:
Service Manual
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR 100e
Item:
DR 100e
Summary:
DR 100e - Solution Service Manual
Edition 1, Revision 10
Changes compared to previous revision:
- Added information about new XF+ 10 / XF+ 14 / XF+ 17 Detector, see chapter 1.3.3.3
- Added reference to Clean installation procedure for DR 100e All-in-One PC: NX 24.00, NX 25.00 and NX 26.00 SU1
- Updated chapter 8.3.1, Clean installation of the NX Workstation
- Updated chapter Release information, see chapters 10.1, 10.2, 10.3 and 10.4
- Added chapter 11.3, Install alternatives for the Aruba AP-303 Access Point
- Updated links and references
- Applied current document template
Publication History:
Published
2025-07-10 10:48:52 CET
Published
2023-06-01 15:40:18 CET
Published
2022-07-01 10:32:22 CET
Published
2021-11-24 15:10:25 CET
Published
2021-08-30 10:19:39 CET
Published
2020-03-11 12:54:31 CET
Published
2019-11-07 15:01:01 CET
Published
2019-03-12 15:59:56 CET
Published
2018-07-25 08:55:24 CET
Published
2018-03-07 10:45:55 CET
Published
2017-09-01 14:46:10 CET
Additional Languages:
No additional languages found

Tplink Access Point Archer T3U AC1300 - Software - driver - Info

Published: Thu Jul 10 10:48:45 2025
Created Date: July 08, 2025
Document Type:
Software
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR Detectors
Item:
Access Points
Summary:
WiFi dongle for DR 100e
Currently the driver for the Wifi dongle Tplink Archer T3U AC1300 is only available via the clean installation image of NX 26.00 SU1.
Publication History:
Published
2025-07-10 10:48:45 CET
Additional Languages:
No additional languages found

DR 100e - SB083 - Release of XD/XF+ Detector in AP mode and Wifi dongle to connect to hospital-network

Published: Thu Jul 10 10:48:11 2025
Created Date: June 30, 2025
Document Type:
Service Bulletin
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR 100e
Item:
DR 100e
Summary:
DR 100e - Service Bulletin No. 83 - Version 1.0
This document contains information about the release of:
- XD/XF+ Detectors configured for use in Access point mode
- In combination with the USB WiFi dongle TP-Link Archer T3U - AC1300 to connect NX to the hospital network for DR 100e digital systems.
Affected systems: XD/XF+ Detector in AP mode and USB Wifi dongle are introduced in production on DR 100e digital systems as of SN 30231 (type 6012/003) and as of SN 40445 (type 6012/004).
Publication History:
Published
2025-07-10 10:48:11 CET
Additional Languages:
No additional languages found

DR wireless networking - Service Manual

Published: Thu Jul 10 10:48:08 2025
Created Date: April 15, 2025
Document Type:
Service Manual
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR Detectors
Item:
DR 10s/DR 14s/Pixium 4343
DX-D 30C/DX-D 35C
DX-D 40/DX-D 45/DX-D 60
DR 10e/DR 14e/DR 17e
XD 10/14/17
XF+ 10/14/17
Summary:
DR wireless networking - Service Manual
Edition 1, Revision 27
Changes compared to previous revision:
- Adapted compatibility matrix in section 1.6: DR 100e now also allows connection of XD and XF+ Detector in access point mode, in combination with an external WiFi adapter tp-link T3U for hospital connection.
- Added step to rename the external WiFi adapter to HOSPITAL, if configured on DR 100e. Refer to page 183.
- Added setup instructions for the USB WiFi adapter tp-link Archer T3U in section 4.17.
- Added troubleshooting for symptom "When tablet / notebook is disconnected from the docking station it takes several minutes until Detector reconnects". Refer to section 7.1.5.
Publication History:
Published
2025-07-10 10:48:08 CET
Published
2025-02-05 09:18:03 CET
Published
2024-06-14 18:01:47 CET
Published
2024-05-17 16:27:38 CET
Published
2024-03-01 16:07:32 CET
Published
2023-11-16 10:54:38 CET
Published
2023-05-19 10:57:58 CET
Published
2023-03-24 13:31:56 CET
Published
2022-10-26 15:36:14 CET
Published
2022-05-16 09:31:52 CET
Published
2021-12-21 15:57:38 CET
Published
2021-11-08 09:43:49 CET
Published
2021-08-19 15:00:06 CET
Published
2021-05-06 15:04:59 CET
Published
2021-02-26 16:07:45 CET
Published
2020-09-07 09:19:25 CET
Published
2020-03-17 17:10:54 CET
Published
2020-03-06 09:35:00 CET
Published
2019-12-05 14:47:40 CET
Published
2019-10-17 16:26:01 CET
Published
2019-07-15 08:41:55 CET
Published
2019-04-03 16:31:29 CET
Published
2019-03-29 16:12:21 CET
Published
2018-06-26 11:34:07 CET
Published
2018-06-05 11:52:31 CET
Published
2018-03-22 09:12:03 CET
Published
2017-10-13 21:13:45 CET
Published
2017-09-01 15:35:41 CET
Published
2017-05-19 15:16:53 CET
Additional Languages:
No additional languages found

XF+ 10 / XF+ 14 / XF+ 17 - Service Manual

Published: Thu Jul 10 08:43:00 2025
Created Date: April 26, 2025
Document Type:
Service Manual
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR Detectors
Item:
XF+ 10/14/17
Summary:
XF+10 / XF+14 / XF+17 - Service Manual
Edition 1, Revision 2
Changes compared to previous revision:
- Added new hardware configuration type for DR 100e with Detector in AP mode. Refer to section 1.10.8.
- Changed recommended IP address for the network interface in NX from 169.254.0.50 to 192.168.0.20. Refer to section 1.15.
- Added additional safety note for Detector in AP mode. Refer to section 2.1.
- Restructured Detector configuration by simplifying configuration overview, removing the Quick Installation Guides and adding section 4.4.7, Performing use case dependent settings.
- Added Detector setup for DR 100e in section 4.4.7.
- Changed setup for Detector in AP mode: In case of PCs with two network interfaces, a fixed IP address in combination with a registry setting to allow multiple network connections is configured. Refer to section 4.4.7.
- Added Troubleshooting section 7.2.5.3, Communication interrupt between Aruba AP and Detector about 7 minutes after boot-up.
- Added Troubleshooting setting 7.2.5.4, With Detector in AP mode, no wireless Detector connection when PC is connected to the hospital network.
Publication History:
Published
2025-07-10 08:43:00 CET
Published
2025-02-05 08:22:42 CET
Additional Languages:
No additional languages found

XD 10 / XD 14 / XD 17 - Service Manual

Published: Thu Jul 10 08:42:47 2025
Created Date: April 26, 2025
Document Type:
Service Manual
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR Detectors
Item:
XD 10/14/17
Summary:
XD 10 / XD 14 / XD 17 - Service Manual
Edition 1, Revision 13
Changes compared to previous revision:
- Added new hardware configuration type for DR 100e with Detector in AP mode. Refer to section 1.8.8.
- Changed recommended IP address for the network interface in NX from 169.254.0.50 to 192.168.0.20. Refer to section 1.13.
- Added additional safety note for Detector in AP mode. Refer to section 2.1.
- Restructured Detector configuration by simplifying configuration overview, removing the Quick Installation Guides and adding section 4.4.7, Performing use case dependent settings.
- Added Detector setup for DR 100e in section 4.4.7.
- Changed setup for Detector in AP mode: In case of PCs with two network interfaces, a fixed IP address in combination with a registry setting to allow multiple network connections is configured. Refer to section 4.4.7.
- Added Troubleshooting section 7.2.5.3, Communication interrupt between Aruba AP and Detector about 7 minutes after boot-up.
- Added Troubleshooting setting 7.2.5.4, With Detector in AP mode, no wireless Detector connection when PC is connected to the hospital network.
Publication History:
Published
2025-07-10 08:42:47 CET
Published
2025-02-05 08:33:15 CET
Published
2024-09-11 09:35:55 CET
Published
2024-05-28 10:40:10 CET
Published
2024-04-24 10:46:21 CET
Published
2024-01-19 08:58:57 CET
Published
2023-12-21 14:57:03 CET
Published
2023-09-13 15:51:59 CET
Published
2022-10-26 15:23:16 CET
Published
2022-06-13 11:45:24 CET
Published
2022-04-19 10:16:59 CET
Published
2021-12-06 12:16:34 CET
Published
2021-08-19 14:28:52 CET
Published
2021-06-07 12:38:28 CET
Additional Languages:
No additional languages found

Application Note - FLFS Stitching Acceptance Test

Published: Wed Jul 09 10:30:02 2025
Created Date: July 09, 2025
Document Type:
Service Application Notes
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR Components
DR Application
Item:
DX-D FLFS Stand
DR FLFS Retrofit
DX-D FLFS Horizontal
Genrad
Summary:
Application Note - FLFS Stitching Acceptance Test
Edition 1, Revision 0
This Application Note instructs the Clinical Application Specialist or Field Service Engineer to perform the FLFS Stitching Acceptance Test.
Whether or not the test has to be performed depends on local regulations.
Publication History:
Published
2025-07-09 10:30:02 CET
Additional Languages:
No additional languages found

Enclosure - DR 100s - B4 CSC Board replacement

Published: Tue Jul 08 11:48:21 2025
Created Date: July 07, 2025
Document Type:
Service Enclosures
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR 100s
Item:
DR 100s
Summary:
Enclosure - DR 100s - B4 CSC board: Replacement
Edition 1, Revision 2
Changes compared to previous revision:
- Section 1 updated.
- Section 5 updated.
- Section 6 added.

This enclosure is part of:
- CENTRAL SYSTEM CONTROL 32 kW, with spare part number: TN+228892/1/1
- CENTRAL SYSTEM CONTROL 40 kW, with spare part number: TN+228892/1/2
It describes all necessary steps for the replacement.
Publication History:
Published
2025-07-08 11:48:21 CET
Published
2025-06-11 11:37:44 CET
Published
2024-04-11 08:18:32 CET
Additional Languages:
No additional languages found

DRYSTAR DT 5.000I B (English-datasheet)

Published: Tue Jul 08 11:15:17 2025
Created Date: July 08, 2025
Document Type:
Data Sheet
Availability:
Internet
Category:
Hardcopy
Language:
English
Subcategory:
Dry Media
Item:
Drystar DT 5.000I B
Summary:
Updated version
Publication History:
Published
2025-07-08 11:15:17 CET
Published
2019-03-11 15:19:48 CET
Published
2018-02-27 09:41:46 CET
Published
2017-08-22 09:26:52 CET
Published
2017-08-11 10:46:21 CET
Published
2017-02-10 08:25:56 CET
Published
2016-10-03 08:42:57 CET
Published
2016-09-26 09:36:04 CET
Published
2015-02-16 12:52:35 CET
Published
2014-04-11 11:57:23 CET
Published
2011-05-24 09:08:25 CET
Published
2010-02-22 02:43:05 CET

DRYSTAR DT 5.000I B (Spanish-datasheet)

Published: Tue Jul 08 11:12:10 2025
Created Date: July 08, 2025
Document Type:
Data Sheet
Availability:
Internet
Category:
Hardcopy
Language:
Spanish
Subcategory:
Dry Media
Item:
Drystar DT 5.000I B
Summary:
Publication History:
Published
2025-07-08 11:12:10 CET
Published
2019-03-11 15:24:53 CET
Published
2018-05-07 10:53:58 CET
Published
2017-08-22 09:51:06 CET
Published
2017-07-31 10:17:17 CET
Published
2016-10-03 08:58:12 CET

DRYSTAR DT 5.000I B (Italian - datasheet)

Published: Tue Jul 08 11:04:07 2025
Created Date: July 08, 2025
Document Type:
Data Sheet
Availability:
Internet
Category:
Hardcopy
Language:
Italian
Subcategory:
Dry Media
Item:
Drystar DT 5.000I B
Summary:
Updated version
Publication History:
Published
2025-07-08 11:04:07 CET
Published
2019-03-11 15:34:20 CET
Published
2018-06-07 11:07:48 CET
Published
2017-08-22 09:43:00 CET
Published
2017-07-31 10:09:09 CET
Published
2016-10-03 09:07:21 CET
Published
2015-02-16 12:56:42 CET
Published
2014-04-11 09:35:42 CET
Published
2011-05-24 09:10:34 CET
Published
2010-02-17 05:11:03 CET

DRYSTAR DT 5.000I B (French-datasheet)

Published: Tue Jul 08 11:03:03 2025
Created Date: July 08, 2025
Document Type:
Data Sheet
Availability:
Internet
Category:
Hardcopy
Language:
French
Subcategory:
Dry Media
Item:
Drystar DT 5.000I B
Summary:
Updated version
Publication History:
Published
2025-07-08 11:03:03 CET
Published
2019-03-11 15:30:13 CET
Published
2018-05-04 09:45:14 CET
Published
2018-03-26 11:30:58 CET
Published
2017-08-22 09:58:16 CET
Published
2017-07-31 10:27:28 CET
Published
2016-10-03 08:49:05 CET
Published
2016-08-05 10:05:06 CET

X-Ray Generator Spellman EDITOR HFe - Service Manual - DR 400/DR 600/DR 501/DR 400C/Valory

Published: Tue Jul 08 08:08:36 2025
Created Date: June 06, 2025
Document Type:
Service Manual
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR Components
Item:
Generator
Summary:
Spellman X-Ray Generator EDITOR HFe - Service Manual
Edition 1, Revision 17
Changes compared to previous revision:
- Updated chapter 01: Service Manual EDITOR HFe X-ray Generators , 06220010_EN_R
- Updated chapter 06: Schematics Diagrams EDITOR HFe, 05-2025
Publication History:
Published
2025-07-08 08:08:36 CET
Published
2022-10-27 10:42:47 CET
Published
2022-10-14 14:36:00 CET
Published
2022-09-15 07:51:49 CET
Published
2022-03-16 09:28:34 CET
Published
2021-12-03 13:14:16 CET
Published
2021-08-27 15:25:24 CET
Published
2021-06-28 10:33:25 CET
Published
2020-12-18 15:21:47 CET
Published
2020-05-28 08:41:03 CET
Published
2019-09-12 10:25:03 CET
Published
2019-07-05 14:29:02 CET
Published
2019-04-17 15:35:37 CET
Published
2017-05-29 15:46:29 CET
Published
2016-06-06 13:58:03 CET
Published
2016-04-20 10:45:35 CET
Published
2016-02-17 08:15:42 CET
Published
2016-02-17 08:12:43 CET
Published
2015-09-22 08:48:30 CET
Published
2015-08-06 10:46:21 CET
Published
2015-04-14 10:28:30 CET
Published
2014-06-30 10:42:44 CET
Additional Languages:
No additional languages found

X-Ray Generator - Software - Sedecal SoftConsole 6.1 (6.01.0100)

Published: Thu Jul 03 08:34:58 2025
Created Date: July 03, 2025
Document Type:
Software
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DR Components
Item:
Generator
Summary:
Generator_SW_Sedecal_SoftConsole_6.01.0100.zip

Software Download via AgfaBox directory (Agfa ID and password required):
Connect to https://agfabox.agfa.net/ and navigate to the file:
Radiology Solutions SW > Imaging_Software > Direct_Radiography > Generator
Publication History:
Published
2025-07-03 08:34:58 CET
Additional Languages:
No additional languages found

DX-D/DR Retrofit System - SB186 - Release of Sedecal SoftConsole 6.1

Published: Thu Jul 03 08:34:56 2025
Created Date: July 02, 2025
Document Type:
Service Bulletin
Availability:
Extranet
Category:
Direct Radiography
Language:
English
Subcategory:
DX-D Retrofit System
Item:
DX-D Retrofit System Generic
Summary:
DR Retrofit - Service Bulletin No. 186 - Version 1.0
- This document announces the release of the Sedecal SoftConsole 6.1.
- It describes the solutions for the mentioned PR cases.

Affected systems:
- Upgrade DX-D 300 and DX-D 600 where the issues occur.
- Upgrade Sedecal Integrated DR Retrofit with SHFR generators where the issues occur.
Publication History:
Published
2025-07-03 08:34:56 CET
Additional Languages:
No additional languages found

Terms of use for the documentation in the library

No part of the user documentation may be reproduced, copied, adapted or transmitted in any form or by any means without the written permission of Agfa N.V.

Agfa N.V. makes no warranties or representation, expressed or implied, with respect to the accuracy, completeness or usefulness of the information contained in the user documentation and specifically disclaims warranties of suitability for any particular purpose. Products and services may not be available for your local area. Please contact your local sales representative for availability information. Agfa N.V. diligently strives to provide as accurate information as possible, but shall not be responsible for any typographical error. Agfa N.V. shall under no circumstances be liable for any damage arising from the use or inability to use any information, apparatus, method or process disclosed in the user documentation. Agfa N.V. reserves the right to make changes to the user documentation without prior notice. The original version of the user documentation is in English.

Copyright Agfa N.V. All rights reserved. Published by Agfa N.V. B-2640 Mortsel – Belgium.